Control menu, Msg’s menu – YSI 5200A User Manual

Page 84

Advertising
background image

YSI 5200A

167

YSI 5200A

166

Notes:

- Always manually poll the 5200A after uploading configuration changes

and verify that the uploaded information is retained. When configuration

changes are uploaded using AquaManager at the same time configuration

changes are being made by a user in the menus at the 5200A, AquaMan-

ager uploads may not be saved depending on the timing of keystrokes

at the 5200A.

- The Run Screen must be displayed during normal operation. If the

5200A is not at the run screen during AquaManager uploading, press-
ing

at the 5200A will only save certain configuration changes.

See Chapter 5 AquaManager - page 172 and AquaManager Online Help for ad-

ditional information.

Control Menu

Press

at the Run Screen to display a list of energized sensor and aux 5200A

system relays. The Control menu identifies the active sensor or aux system and

control and/or alarm condition. See Sensor Setup Controls and Alarms - page 132

for additional information.

Notes:

- Press

to display time stamped detail of control and alarm relay(s).

- Alarm message appears at the Control menu for aux and sensor systems

when alarms are enabled regardless if an alarm relay is assigned.

- Active timer (10 event timer and feed timer) systems are not displayed

at control menu.

- General alarm systems are not displayed at the Control menu regardless

if general alarm relay is enabled.

- Use

to scroll list when scroll bar appears at menu.

- Conductivity and pH Control Range Low message remains at the

Configuring the 5200A

menu when the conductivity or pH system is disabled and a control low

relay is energized. Power cycle the 5200A to clear the control low range

message at the Msg’s menu.

When no 5200A control or alarm system(s) are active, the below screen is dis-

played.

Msg’s Menu

Calibration, Sensor, and Event logs contain 5200A system information. The 5200A

Sensor log stores 1000 records. The Calibration log stores 500 records. The Event

log stores 1250 records. Once the log memory is full, the oldest records are over-

written with more recent records. The number shown in ( ) at the Msg’s menu

reflects the total number of records in the log. The 25 most recent records for each
log are displayed at the

menu.

Use

to scroll and press

to view

time stamped detail record.

Use AquaManager to capture all records from 5200A logs and store them to a PC

database. Polling times should be determined by logging intervals. Example - if the

sensor log is recording values every minute and the sensor log holds 1000 records,

polling should occur at least every 27 hours. This would allow for all records to be

saved to the PC before being overwritten on the 5200A. Set the data log interval so

that analyzing data based on the configured interval will be helpful. Poll at intervals

that don’t result in gaps in data.

Configuring the 5200A