- Configuring a 5200A as a Master overrides any slave configuration that

had been previously saved.

- Slave alarms (slave alarm at master and 5200A alarm emails) will not oc-

cur if slave node serial number is not recognized at master node at steps

10. A or 10. B above.

Termination Relay

Menus → System → Communications → Network →Termination Relay

The termination relay must be enabled on the first and last 5200A on a network. All

other devices should have the termination relay disabled.

Send Test Email Msg

Menus → System → Communications → Send Test Email Msg

Send a test email to verify proper configuration of ethernet menus. Test emails are

only transmitted from configured master 5200As.

Note:

- Test emails are sent to all configured email address.

- Exit to Run Screen is required before the test email is sent.

Example of a 5200A test email sent from an ethernet device:

Subject Line: 02/04/10 03:50:05 PM YSI5200A Email Alarm Test Message

From: [email protected]

Sent: Wed, Feb 04, 2010 3:49 pm

To: [email protected]

CC: [email protected]

Configuring the 5200A

Email Retry Delay

Menus → System → Communications → Email Retry Delay

Configure the Email Retry Delay for the time interval that emails should be retrans-

mitted after the initial alarm email is sent. The initial alarm email is sent immediately

after the general alarm hold off time or after the aux hold off time. Emails will be

retransmitted until one of the bottom front panel softkeys is pressed, regardless if

the alarm condition still exists.

Note:

- On a network, emails are retransmitted according to the slave retry delay

configuration of the slave in alarm and not the master’s email retry delay

configuration. Please note, a slave Email Retry Delay menu will display

(Not available) because of its slave configuration; however, this feature

is available and has a default value of 15 minutes that can be configured

by the user.

- The Retry Delay should be configured at intervals ≥ 15 minutes.

- AquaManager alarm emails are retransmitted based on the 5200A slave(s)

and master email retry delay configuration.

Baud Rate RS232

Menus → System → Communications → Baud Rate

Configure the Baud rate (for RS232 connections only) in the 5200A to match the

baud rate configured in AquaManager. Use the default rate of 115200 for fastest

communication speed.

1. Enter Network menu

Menus → System → Communications → Network →Baud Rate

2.

Press

or

to scroll and highlight a selection.

3.

Press

to confirm selection.

4. Press

to save configuration and return to Run Screen.