Softkey label

Function
Exit menu and return to Run Screen saving

any menu configuration changes.
Exit menu (to previous menu) without sav-

ing configuration changes.
Back to previous menu. Changes to configu-

rations are saved at return to Run Screen.
Prompt to take next action, i.e. enter a value

at calibration menus.
Save alpha and numeric configurations and

calibrations.

Disable/Enable menu option, i.e. buzzer.

Scroll left and right in alpha and numeric

keypad screens.
To acknowledge and save sensor calibration.

Prompt to confirm “Are you sure?” message

when performing certain functions, i.e. reset

to factory and clear data log menus.
Digital input control configuration from

Sensor Setup Menu.

Table 4.1

Run Screen

The Run Screen displays measurements and inputs from enabled sensors and condi-

tions - figure 4.2. At power up, the Run Screen is displayed immediately after the

YSI splash screen (the YSI splash screen appears for approximately 2 seconds). Icons

displayed at the Run Screen represent individual and systemwide conditions. See

Run Screen Icons - page 70.

Run Screen and sensor display formats are user selectable. See Display Settings -

page 111 and Sensor Setup Menus - page 131.

Notes:

- The Run Screen must be displayed during normal operation. All alarm

functions (5200A alarm emails, energizing of enabled alarm relays,

alarm icons, and buzzer) are suspended when in 5200A instrument

menus. Alarm systems are reset when menus are exited to Run Screen.

Configuring the 5200A

- Always manually poll the 5200A after uploading configuration changes

and verify the uploaded information is retained. When configuration

changes are uploaded using AquaManager at the same time configuration

changes are being made at the instrument, the AquaManager uploads

may not be saved depending on the timing of keystrokes at the 5200A.

- The Run Screen must be displayed during normal operation. If the

5200A is not at the Run Screen during AquaManager uploading, press-
ing

at the 5200A will only save certain configuration changes.

- At power up, 5200A systems have a 30 second hold off allowing sensor

values to stabilize before enabling any control and/or alarm devices.

Figure 4.2

Configuring the 5200A