Calibration log, Conductivity, Directions to calibrate conductivity sensor – YSI 5200A User Manual

Page 39

Advertising
background image

YSI 5200A

77

YSI 5200A

76

Figure 4.5

Notes:

- Insert port plugs into unused sensor port(s) - figure 4.5.

- The sensor must be enabled to calibrate.

- If you use another container for calibration be sure to handle the sensors

will care to protect them from possible physical damage. (See YSI 5562

probe sensor guard - figure 3.12 on page 36 or YSI 5561 probe sensor

guard - figure 3.11 on page 35.) Use a ring stand and clamp to secure

the probe assembly.

- Older sensors may take longer to calibrate than new ones because the sensor

readings may take longer to stabilize.

Configuring the 5200A

Calibration Log

Calibration data is stored in the 5200A calibration log. A calibration data record

includes the type of sensor calibrated, the date and time of the calibration and the
calibration method. Press

at the Run Screen to view the calibration log

- page 168.

Notes:

- Use AquaManager to view and save calibration data on a PC. In addition

to the calibration log information listed above, AquaManager calibration

logs also include user inputs 1 and 2 and the pre and post calibration slopes/

offsets of the sensor(s) when applicable.

- Information from cancelled calibrations is not stored.

Conductivity

Menus → Calibration→ Conductivity

See figure 4.5 to determine which sensor is the conductivity sensor. The conductivity

sensor calibration is a one-point calibration. Calibrate using a traceable conductiv-

ity solution that has a value close to the conductivity value of the water system to

be monitored.

Directions to Calibrate Conductivity Sensor

1. Connect probe/cable assembly - page 61.

2. Read Getting Ready to Calibrate - page 75.

3. Fill calibration/storage cup with 55 ml of traceable conductivity calibration

solution that has a known value.

4. Carefully immerse the sensors into solution. Ensure that the top vent

hole of the conductivity sensor is fully immersed in the solution - figure

4.5, page 76.

5. Gently move the probe module up and down to remove any air bubbles

trapped in the conductivity sensor vent hole.

6. Screw the calibration/storage cup onto the threads of the probe module.

7. Enter conductivity calibration menu:

Menus → Calibration→ Conductivity

8. Allow at least one minute for the temperature to equilibrate.

Configuring the 5200A

Advertising