Timed and normal control, Directions to configure ph sensor, Temperature – YSI 5200A User Manual

Page 76: Range low and high control

Advertising
background image

YSI 5200A

151

YSI 5200A

150

Notes:

- To configure a negative set point, control, and alarm values enter “-” after,

not before, the numeric digit(s).

- See 30 Second Sensor Hold Off - page 141.

- Only whole numbers can configured for ORP set point, control and

alarm values.

pH

Menus →Sensor Setup→pH

Read Sensor Setup Control and Alarms - page 132 before configuring pH system.

Timed and Normal Control

pH high and low control can operate in two different modes, normal or timed. In the

normal mode, control relays are energized until the set point value is reached. The

normal mode can be used for either low control, high control, or both. For normal

control, configure dose up/dose down to 0. The dwell time will be ignored.

In the timed mode, an energized control relay is energized for the dose time, re-

gardless if the sensor value has returned to the set point value. After the dose time

expires, the control relay de-energizes for the configured dwell time. After the dwell

time expires and if the value is still in control range, the control relay will reenergize.

A dwell time must be configured in the timed mode or the control relay will never

de-energize regardless of sensor value.

Set Point

Control

Mode only

Configuring the 5200A

Directions to Configure pH sensor

1. Connect probe assembly - page 61.

2. Enter pH Sensor Setup menu

Menus →Sensor Setup→pH

3.

Use

and

to scroll and highlight submenus.

4.

Press

to select.

5. Configure submenus.
6. Press

to save configuration and return to Run Screen.

Notes:

- pH set point, control, and alarm values can be up to five (5) characters,

including the decimal point. Values truncate to 100ths when no 10s

digit is configured.

- See 30 Second Sensor Hold Off - page 141.

- Active dosing timers are reset after systemwide events. Systemwide events

de-energize relays. Example - pH dose up relay is configured for two (2)

minutes. Energized dose up relay de-energizes when 5200A goes into Svr

Hold. The dose up relay had been energized for 15 seconds before the

Svr Hold occurs. Svr Hold time is one (1) minute. After Svr Hold time

expires, the control system is reset. If the sensor is reporting a control

low condition, the dose on time will be for two (2) minutes and not the

remaining 45 seconds of the dose up time prior to Svr Hold.

- pH Control Low Range message remains at the

menu when the

pH system is disabled and when a control low relay is energized. Power

cycle the 5200A to clear the control low range message at the Msg’s menu.

Temperature

Menus →Sensor Setup→Temperature

Read Sensor Setup Control and Alarms - page 132 before configuring Temperature

system. Since most heaters and chillers have their own thermostats, their thermo-

stats can be used as a backup temperature control system. Set the heater thermostat

several degrees above the desired temperature and the chiller thermostat several

degrees below the desired temperature.

Range Low and High Control

Temperature high control can operate in two different modes, normal or timed.

Temperature low control operates in normal mode only. In the normal mode, control

relay is energized until the set point value is reached. For normal control configure,

set chiller on time to 0. The chiller off time will be ignored.

Set Point

Control

Mode only

Configuring the 5200A