3 storage, 4 location of installation and ambient conditions, Caution – BINDER KBF 115 User Manual

Page 24

Advertising
background image

KBF / KBF P (E5.3) 09/2014

page 24/110

3.3 Storage

Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-

rary decommissioning (chap. 17.2).
Permissible ambient temperature range during storage:

If the steam humidifying system has NOT been emptied: +3 °C / 37.4 °F to +60 °C / 140 °F.

After BINDER Service has emptied the steam humidifying system: -10 °C / 14 °F to +60 °C / 140 °F

With temperatures below +3 °C / 37.4 °F, water must be completely removed from the humidifying sys-

tem.

CAUTION

Storage below +3 °C / 37.4 °F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.

Contact BINDER Service before any transportation below +3 °C / 37.4 °F.

Permissible ambient humidity: max. 70 % r.H., non-condensing

CAUTION

Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H.

for a long period.

Dry the appliance completely before shut-down:

Set the humidity to 0 % r.H. and turn on humidity switch (4).

Set the temperature set point to 60 °C / 140 °F for approx. 2 hours (Manual mode).

Only then, shut down the unit at the main power switch (3) and close the tap of the

water supply.

When after storage in a cold location you transfer the unit to its warmer installation site, condensation

may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and

is completely dry.
In case of a prolonged temporal decommissioning: Leave the unit door open or remove the access port

plugs.

3.4 Location of installation and ambient conditions

Set up the constant climate chamber on a flat, even surface, and in a well-ventilated, dry location and

align it using a spirit level. The site of installation must be capable of supporting the unit’s weight (see

technical data, chap. 19.4). The chambers are designed for setting up inside a building (indoor use).

CAUTION

Danger of overheating.
Damage to the unit.

Do NOT set up unit in non-ventilated recesses.

Ensure sufficient ventilation for dispersal of the heat.