10.3.2 Temperature safety device class 3.2

The safety device class 3.2 is equivalently set to a minimum temperature the

unit will not fall below. This protection against prohibited low temperatures

can, for example, serve to protect sensitive cultures from cooling down too

much.
If the control knob (15) is turned to its minimum (position 1), the safety de-

vice class 3.2 has no effect. If it is set to a temperature somewhat lower than

that selected by means of the controller, it functions as a protective device

for the material under treatment.

If the temperature safety device class 3.2 has assumed regulation, identifiable by the red alarm lamp

(15a) lighting up, the message “TEMP ALARM” on the controller display and the buzzer sounds, please

proceed as follows:

Reset the buzzer with the “RESET” key on the controller.

Disconnect the unit from the power supply.

Have an expert examine and rectify the cause of the fault.

Start up the unit again as described in chap. 5.

Setting:
To check the response temperature of the safety device class 3.2, put the unit into operation and set the

desired set point at the temperature controller.
The sections of the scale from 1 to 10 correspond to the temperature range from -40 °C / -40 °F to +160

°C / 320 °F and serves as a setting aid.

Turn the control knob (15) of the safety device by means of a coin to position

1 (thermostat without effect).

When the set point is reached, reset the safety device to its trip point (turn it

clockwise).

The trip point is identifiable by the red alarm lamp (15a), the message “TEMP

ALARM” on the controller display and the buzzer sounds. Reset the buzzer

with the “RESET” key on the controller.

The optimum setting for the safety device is obtained by turning the control

knob counter-clockwise by approximately two scale divisions, which shuts off

the red alarm lamp (15a).

Figure 20: Setting

safety device class 3.2

Check the setting regularly and adjust it following changes of the set-point or charge.

Function check:
Check the temperature safety device class 3.2 at appropriate intervals for its functionality. It is recom-

mended that the authorized operating personnel should perform such a check, e.g., before starting a

longer work procedure.