Warning, Caution – BINDER KBF 115 User Manual

Page 77

Advertising
background image

KBF / KBF P (E5.3) 09/2014

page 77/110

When using the communication software APT-COM™ 3 DataControlSystem (option, chap.

15.1) via the Ethernet interface of the constant climate chamber for data acquisition, the alarm

is not automatically transmitted to the APT-COM™ protocol.

Set the tolerance limits for recording limit value excesses separately in APT-COM™ 3.

15.6 Water protected internal socket (option – not for UL units)

The internal socket is splash proof.
IP system of protection 67 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 90 °C / 194 °F.

WARNING

Exceeding the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.

Do NOT exceed the temperature set-point of 90 °C / 194 °F.

Set the safety controller to 90 °C / 194 °F.

With optional temperature safety device, class 3.3, set the mechanical thermostat class

3.1 to 90 °C / 194 °F.

Heat emission of electrical devices connected inside the chamber may modify the temperature

and humidity range.

CAUTION

Risk of short circuit.
Damage to the unit.

Use the supplied plug only (IP protection type 67). Plug it in and tighten it by screwing

to secure contact.

If the socket is not used, close the screw lid and turn it to secure.