BINDER KBF 115 User Manual

Page 80

Advertising
background image

KBF / KBF P (E5.3) 09/2014

page 80/110


(2) Cable connections

Connect the plug of the cable to the socket (22) at the rear of the unit.
The socket (22) is marked with a sticker:


Socket (22)








(17)

(18)

Figure 31: Connections at the unit rear KBF / KBF P 240


(3) Hose connections

Plug the freshwater hose into the hose connection (21) above the freshwater can and secure it with

a hose clamp. You can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwa-

ter connection “IN” (18) at the rear of the unit.

When the freshwater can is empty, within 60 seconds the alarm message “HUMID ALARM” will be dis-

played on the controller, the buzzer sounds (chap. 11), and the humidity system turns off.

To guarantee humidification during 24 hours even at high humidity set-points with manual

water supply, we recommend filling the freshwater can (option) at the end of each day.