KBF / KBF P (E5.3) 09/2014

page 50/110

9.4 Selecting between “set-point ramp” and “set-point step”

Set-points always refer to the start of a program section, i.e., at the beginning of each program section

the entered “set-point” is targeted. During program section operation, the temperature or humidity gradu-

ally passes to the set-point entered for the next section.
By appropriate planning of the program section timing, you can enter all kinds of temperature and humidi-

ty transitions:

Gradual temperature / humidity changes “set-point ramp”
The set-point changes its value gradually while proceeding from one program section to the next one

during the programmed section length. The actual temperature or humidity value (X) follows the con-

tinually moving set-point (W) at any time.

Program sections with constant temperature / humidity
The initial values of two subsequent program sections are identical; so the temperature or humidity

remains constant during the whole time of the first program section.

Sudden temperature / humidity changes “set-point step”
Steps are temperature or humidity changes (ramps) that occur during a very short interval. A section

with a different set-point follows two program sections with an identical set-point. If the duration of this

transitional program section is very short (minimum entry 1 sec), the temperature or humidity change

will proceed rapidly within the minimum amount of time.

Figure 17: Possible temperature or humidity transitions


The following chapter offers examples of programming a “set-point ramp” and a “set-point step”.

01

02

03

04

05

06

07

08

09

W

t