BINDER KBF 115 User Manual

Page 48

Advertising
background image

KBF / KBF P (E5.3) 09/2014

page 48/110


Enter the individual values of the selected program section.

Program editor Abs.Nr. 6

****.*

Setpoint 1

+100.0

FAN

Operating contacts

Time

00:45:00

Repeat Section

5

Repeat Number

10

Tol.-band min.

-1999.0

Tol.-band max.

+9999.0

Parameter set

1

Setpoint 1

Temperature value at the start of the program section

FAN

Fan speed in %

Operating contacts Operation lines (no function)
Time

Duration of the program section

Repeat Section

No. of start section in case of repeat cycles

Repeat Number

No. of duplicates in case of repeat cycles

Tol.-band min.

Temperature limits (maximum / minimum temperature)

(In case of exceeding: temporary program stop)

Tol.-band max.
Parameter set

Pre-selected value (Do NOT change!)

Select the parameters via the arrow keys and confirm by pressing “ENTER”.
Then enter the values via the arrow keys, and confirm by pressing “ENTER”.

For a negative set-point entry, enter the numerical value first, and then the minus sign (-).

With set-point type “

Limit”, the user shall adapt the safety controller (chap. 10.2) or the tem-

perature safety device class 3.3 (option, chap. 10.3) to the highest temperature set-point value

of the program actually used. Check the safety device for each temperature program and

change it if necessary. Set the safety controller set-point or the set-point of the temperature

safety device class 3.3 (option) by approx. 2 °C to 5 °C above the controller temperature set-

point.
Recommended setting: Set-point type “

Offset” with set-point 2 °C.

Entering the set point values for KBF P when operating without illumination:
Adapt the setting of the temperature set-point according to the table in chap. 8.2.

Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber

approximates ambient temperature. The fan turns at 50% rate (factory setting).
Note for KBF P: When the illumination is operating, the inner chamber temperature may rise considera-

bly after the program ends. When operating with illumination, an additional program section at the end of

the program is therefore required. This program section gives the desired final temperature and is re-

peated infinitely (enter the number of cycles “Cy” as “-1“), i.e., until cancelling the program manually

(press the EXIT or AUTOMATIC button).