4.5 Installation and connection of the light cassettes (KBF P)

You can insert the light cassettes in

different heights onto the beads at the

lateral walls of the unit. Insert and pull

out the light cassettes only by the han-

dles.
Connect the cables of the light cas-

settes to its closest connection socket

at the right side in the back of the

chamber.
Put the waterproof plug on the connec-

tion socket. When the plug has en-

gaged, turn the locking-nut of the plug

several times to the right up to its final

stopping point. The plug is now auto-

matically sealed into the socket.

Figure 14: Connecting the light cassettes

Use the covers supplied to protect any unused connection socket.

Do not place any charging material directly onto the light cassettes because those are heated by the fluo-

rescent tubes which would lead to exposing the charging material to undefined temperatures. The tem-

perature directly below or on the light cassettes is not equal to the temperature displayed at the tempera-

ture controller.

Place the charging material on the supplied racks below the light cassettes.

CAUTION

The light cassettes will become hot with temperature set-points >40 °C.
Danger of burning.

Do NOT touch the light cassettes during operation.

Let the light cassettes cool down before changing their position.

When operating the chamber at temperatures > 60 °C / 140 °F, remove the light cassettes.

Otherwise, the life expectancy of the fluorescent tubes will be considerably reduced.