Danger – BINDER KBF 115 User Manual

Page 76

Advertising
background image

KBF / KBF P (E5.3) 09/2014

page 76/110

15.4 Analog outputs for temperature and humidity (option)

With this option the chamber is equipped with analog outputs 4-20 mA for temperature and humidity.

These outputs allow transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (6) in the right lateral control panel as follows:

ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –
PIN 2: Temperature +
PIN 3: Humidity –
PIN 4: Humidity +
Humidity range: 0 % r.H. to 100 % r.H.
Temperature range: -10 °C / 14 °F to +100 °C / 212 °F
A suitable DIN plug is enclosed.

Figure 25: Pin allocation of DIN socket (6) for option analog outputs

15.5 Zero-voltage relay alarm outputs for temperature and humidity (option)

The chamber equipment with optional zero-voltage relay outputs for temperature

and humidity (option) permits the transmission of alarms to a central monitoring

system. Connection is established via a DIN socket (7) located on the right lateral

control panel.

Figure 26: Pin configuration of the DIN socket (7)

Temperature contact

Humidity contact

Pin 1: Pin
Pin 2: Make

Pin 3: Pin
Pin 4: Make

In case of a temperature alarm, pins 1 and 2 are open; with humidity alarm, pins 3 and 4 are open. This

happens simultaneously with the alarm message shown on the controller display.
In case of power failure, both contacts are open.
Maximum loading capacity of the switching contacts: 24V AC/DC - 2,5A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.5A.

Do NOT connect any devices with a higher loading capacity.

A temperature and humidity alarm message will remain visible on the controller display during the whole

time of the alarm transmission via the zero-voltage relay outputs.
As soon as the cause of the alarm is rectified, you can reset the alarm transmission via the zero-voltage

relay outputs together with the alarm message at the controller display by pressing the “RESET” key.
In case of power failure, transmission of the alarm via zero-voltage relay outputs remains active for the

duration of the power failure. Afterwards, both contacts will close automatically.

3

4

1
2