4.4 Electrical connection

The constant climate chamber comes with a fixed power connection cable that has a length of 1800 mm /

5.9 ft. The socket must also provide a protective conductor.

Model

Power plug

Voltage

+/-10 %

Current type

Power

frequency

Unit fuse

KBF 115
KBF 240
KBF 720
KBF P 240
KBF P 720

Shock-proof

plug

200 V to 240 V

1N~

50/60 Hz

16 Amp

KBF 115-UL
KBF 240-UL
KBF 720-UL
KBF P 240-UL
KBF P 720-UL

NEMA 6-20P 200 V to 240 V

2~

50/60 Hz

16 Amp

Prior to connection and start-up, check the power supply voltage. Compare the values to the specified

data located on the unit’s type plate (left unit side, bottom right-hand, see chap. 1.4).
When connecting, please observe the regulations specified by the local electricity supply company and

as well as the VDE directives (for Germany). We recommend the use of a residual current circuit breaker.

Pollution degree (acc. to IEC 61010-1): 2

Installation category (acc. to IEC 61010-1): II

CAUTION

Danger of incorrect power supply voltage.
Damage to the equipment.

Check the power supply voltage before connection and start-up.

Compare the power supply voltage with the data indicated on the type plate.

See also electrical data (chap. 19.4).

To completely separate the unit from the power supply, you must disconnect the power plug.

Install the unit in a way that the power plug is easily accessible and can be easily pulled in

case of danger.

Remark when operating with a power frequency of 60 Hz:

WARNING

High leakage current.
Electrical hazard.

Earth connection essential before connecting supply. Check socket before inserting

plug.

When connected to a power supply 1N~ with a frequency of 60 Hz, a leakage current of more than 3.5

mAmp is possible. If grounding through the power cable is insufficient or missing, the leakage current

may flow through the user’s body. Correct installation of the professional grade power socket provided by

the user safely avoids this. Before connecting the unit to the socket, please check its grounding contact

type plug for appropriate construction and if it is undamaged.