12.1 Function of the humidifying and dehumidifying system

Humidifying system
The humidifying system is located in the humidity generation module. In a cylindrical container with a

volume of about 2 liters an electrical resistance heating evaporates water. The water content is kept ex-

actly at the boiling point, and thus steam can be immediately generated in sufficient quantity for rapid

humidity increases or for compensation of humidity losses, e.g. by door openings. Condensation forming

on the outer walls of the useable volume is led through a water drain in the outer chamber into the

wastewater can which is pumped off automatically to the wastewater pipe when required.
Freshwater
You can supply the unit with freshwater via a water pipe or by manually filling a freshwater can (option,

chap. 15.9) You can mount the can on the rear of the unit or place it next to the unit.

In order to ensure accurate humidifying, observe the following points with regard to the

freshwater supply:

Supply pressure 1 to 10 bar / 14.5 to 145 PSI when connecting to a water pipe

Water type: deionized (demineralized) water

To ensure humidification during 24 hours even at high humidity set-points with manual

water supply, we recommend filling the freshwater can (option) at the end of each day.

Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F.

BINDER GmbH is NOT responsible for the water quality provided by the customer.
Any problems and malfunctions that might arise following use of water of deviating quality is

excluded from liability by BINDER GmbH.

Automatic freshwater supply via water pipe
With this type of supply, the humidity system is continuously functional.
Manual freshwater supply via freshwater can (option, chap. 15.9)
With this type of supply, the humidity system is functional only if the water can is sufficiently filled. Check

the filling level daily. The water reserve in the can is sufficient for a period, which may last between one

and several days, depending on the humidity demand (entered humidity set-point and number of door

openings).
Wastewater
The condensation water from the interior is collected in an internal can with a volume of approx. 0.5 liters.

It is pumped off via the wastewater pipe.
Dehumidifying system
When humidity switch “ON / OFF” (4) (located on the right lateral control panel) is “ON”, the constant

climate chamber KBF dehumidifies as needed in order to reach the entered humidity set-point” inside the

control range of temperature and relative humidity (Figure 21).
Dehumidification occurs in case of need by means of defined dew point undershoot of several evapora-

tors of the refrigeration system. The condensate which forms is carried away as wastewater.
If the humidity system is turned off while there are descending temperature curves, then operation of the

refrigeration system may cause dehumidification of the charging material.
For error indications concerning water supply and humidity system, see chap. 11.1 and 18.