background image

KBF / KBF P (E5.3) 09/2014

page 35/110

5.4 Turning on the unit

Set the main power switch (3) to position I. The pilot lamp shows the unit is ready for operation.

Observe a delay time of about 30s between turning Off and On again. Otherwise an initializa-

tion problem may occur (display showing e.g. “–1999”).

Note that the chamber is in stand-by mode when the main power switch is in position I and the controller

display is dark. Turn on the unit by pressing any button. When turned on, the unit functions in the operat-

ing mode entered before turning off. In Manual Mode (HAND), the controller regulates temperature and

humidity to the last entered set-points, and in Program Mode (AUTO) it regulates temperature and humid-

ity to their set-points reached during the previous program operation.


Structure of toggling between Idle Mode / Manual Mode / Program Mode:

Idle Mode

HAND

EXIT

Menu program start

Manual Mode (HAND)

Program Mode (AUTO)

or

EXIT