Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 102

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

7-4

SB Temp Hi (Sample block heater temperature high, > 190ºC)

Probable Cause

Recommended Corrective Action

High noise in OCX power supply

Check power supply for line noise or voltage fluctuations. Install power line filter kit
(PN 6A00171G01) or high quality line filter for input power.

O2 Temp Very Hi (Oxygen sensor heater over maximum temperature, > 820ºC)
O2 Htr Rmp Rate (Oxygen sensor heater over maximum temperature ramp rate)

Probable Cause

Recommended Corrective Action

Incorrect O

2

heater wiring

Check O

2

heater wiring per Figure 7-3 and Figure 6-6. Check the wiring at the heater

and inside the electronics housing. Correct wiring fault. Perform Reset procedure in
Section 3, Configuration and Startup to continue operation.

Incorrect O

2

thermocouple wiring

Check O

2

thermocouple wiring per Figure 7-3 and Figure 6-6. Check the wiring at the

thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.

Electronics package failure

Replace electronics package.

COe Temp Very Hi (Combustion sensor heater over maximum temperature, > 400ºC)
COe Htr Rmp Rate (Combustibles sensor heater over maximum temperature ramp rate)

Probable Cause

Recommended Corrective Action

Incorrect COe heater wiring

Check COe heater wiring per Figure 7-3 and Figure 6-7. Check the wiring at the heater
and inside the electronics housing. Correct wiring fault. Perform Reset procedure in
Section 3, Configuration and Startup to continue operation.

Incorrect COe thermocouple wiring

Check COe thermocouple wiring per Figure 7-3 and Figure 6-7. Check the wiring at the
thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.

Electronics package failure

Replace electronics package.

SB Temp Very Hi (Sample block heater over maximum temperature, > 260ºC)
SB Htr Rmp Rate (Sample block heater over maximum temperature ramp rate)

Probable Cause

Recommended Corrective Action

Incorrect sample block heater wiring

Check sample block heater wiring per Figure 7-3 and Figure 6-6. Check the wiring at
the heater and inside the electronics housing. Correct wiring fault. Perform Reset
procedure in Section 3, Configuration and Startup to continue operation.

Incorrect sample block thermocouple wiring

Check sample block thermocouple wiring per Figure 7-3 and Figure 6-6. Check the
wiring at the thermocouple and inside the electronics housing. Correct wiring fault.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

Electronics package failure

Replace electronics package.

O2 TC Open (Oxygen sensor heater thermocouple open)

Probable Cause

Recommended Corrective Action

O

2

thermocouple or thermocouple circuit open

Check O

2

thermocouple and circuit wires for breaks or loose connections per Figure 7-3

and Figure 6-6. Repair breaks or loose connections or replace failed thermocouple.
Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

COe TC Open (Combustibles sensor heater thermocouple open)

Probable Cause

Recommended Corrective Action

Combustibles sensor thermocouple open

Check thermocouple resistance (lead to ground) of reference and active thermocouples
per Figure 7-3 and Figure 6-7. If either thermocouple is open or shorted to ground,
replace combustibles sensor.

Table continued on next page