Ocx 8800, Instruction manual, Alarm relay events – Emerson ROSEMOUNT OCX 8800 User Manual

Page 109

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

7-11

OCX 8800

ALARM RELAY EVENTS

The OCX 8800 contains an alarm relay that can be configured to activate on
one of twelve different groups of events. These event groups, and the
conditions that trigger them, are listed in Table 7-2, Alarm Relay Event
Groups.

Table 7-2. Alarm Relay

Event Groups

Alarm Relay Event

Alarms/Conditions

In Calibration

Calibration in progress

O

2

Temperature Error

Oxygen sensor heater temperature low (O2 Temp Low)
Oxygen sensor heater temperature high (O2 Temp Hi,

O2 Temp Very Hi)

RTD excitation current error (Ref Curr Err)

Heater Failure

Oxygen sensor heater could not reach final temperature

(O2 HTR Failure)

Combustibles sensor heater could not reach final temperature

(COe Htr Failure)

Sample Block sensor heater could not reach final temperature

(SB Htr Failure)

O

2

Sensor Error

Oxygen sensor resistance high (O2 Sensor R High)
Oxygen sensor disconnected (O2 Sensor Open)

Calibration Failure

Calibration failed (Cal Failed)

Calibration Warning

Calibration warning (Cal Warning)

Board Temperature High

Electronics temperature maximum exceeded (Board Temp Hi)

Unit Failure

Any non-recoverable or heater relay off alarm (O2 Temp Hi,

O2 Temp Very Hi, COe Temp Hi, COE Temp Very Hi,
SB Temp Hi, SB Temp Very Hi, O2 Htr Ramp Rate,
COe Htr Rmp Rate, SB Htr Ramp Rate, O2 TC Shorted,
O2 TC Reversed, COe TC Shorted, COe TC Reversed,
SB TC Shorted, SB TC Reversed, ADC Failure,
ADC Ref Error, Board Temp Hi, EEPRM Chksm Fail,
Line Freq Error, Line Voltage Low, Line Voltage Hi,
Htr Relay Failed, Out Brd Fail)

Sample Block

Temperature Error

Sample block heater temperature low (SB Temp Low)
Sample block heater temperature high (SB Temp Hi,

SB Temp Very Hi)

RTD excitation current error (Ref Curr Err)

COe Sensor

Temperature Error

Combustibles sensor heater temperature low (COe Temp Low)
Combustibles sensor heater temperature high (COe Temp Hi,

COe Temp Very Hi)

RTD excitation current error (Ref Curr Err)

Power Input Error

AC power line frequency out of usable range (Line Freq Err)
AC power line voltage below minimum (Line Voltage Low)
AC power line voltage above maximum (Line Voltage Hi)

All Alarms

Any alarm