6-32

Figure 6-25. Electronics Housing

Cable Connections

37390016

RED
BLK

WHT

GRN

BLK

BLK

BLU
BLK

YEL

BRN

RED

BLK

BLK

ORG

BLK

WHT

T/C CO+
T/C CO-

T/C SB+
T/C SB-
T/C O2+
T/C O2-

O2 CELL+
O2 CELL-

SHIELD

To

Ground

Screw

GRN

ELECTRONICS HOUSING

O2 Cell

& Thermocouple

Connector (J5)

COe Sensor

& CJC

Connector (J4)

EXC+
CO ACT+
CO ACT-

CO REF+
CO REF-
CJC+
CJC-
EXC-

Heater
Power Cable

Signal Cable

NOTE: Wire colors shown are for cables supplied

by Emerson Process Management.

1

#1

#1

#1

#1

#1

To Ground

Screw

SHLD

Heater Power

Connector (J3)

#1

2HTR SB

1HTR O

2

1HTR SB

2HTR CO

1HTR CO

2HTR O

2

WHT

ORG

BLU

BLK

YEL

RED

Advertising