Install Sensor Housing

1. Connect the test gas, reference air, eductor air, and dilution air lines to

the sensor housing.

2. Remove the sensor housing cover.
3. If removed, install the power and signal cables and the customer power

and signal conduits and wiring at the sensor housing.

4. Connect the signal cable to O

2

and T/C terminal blocks and to the CO

and CJC terminal blocks, Figure 6-2. Connect the heater power cable to
the HTR terminal blocks.

5. Refer to Figure 6-1 and make sure all test gas lines and electrical

connectors are complete.

6. Install the sensor housing cover.
7. Restore power to the system. Allow OCX to reach normal operating

temperature before installing in a hot process stack.

8. Insert and bolt the sensor housing in the stack and install insulation.
9. Turn on the test gases at the cylinders and open the instrument air

supply valve.

Remove Remote Electronics Housing

1. Turn off power to the system.
2. Shut off the test gases at the cylinders and close the instrument air

supply valve, Figure 6-1.

3. Disconnect the test gas and instrument air lines from the remote

electronics housing.

4. Remove the cover from the electronics housing to expose the

electronics housing terminal blocks, Figure 6-3.

5. Disconnect and remove the power leads from the AC power input

terminal block. Remove the ground lead from the ground stud.

6. Disconnect and remove the O

2

and COe signal leads from the 4-20 mA

signal output terminal block.

7. If used, disconnect and remove the external relay leads from the alarm

output relay terminal block.

8. Disconnect the signal cable leads from O

2

cell and thermocouple

connector (J4), and from COe and CJC connector (J5), Figure 6-4.

9. Disconnect the heater cable leads from the heater power connector

(J3).

10. If moving the electronics housing to another work site, disconnect and

remove the power and signal cables and customer wiring conduits from
the housing.

11. Remove the remote electronics housing from its mounting and move it

to a suitable work area.