Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 105

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

7-7

OCX 8800

Figure 7-2. Oxygen and Cell

Output

Cal Warning (Calibration warning)
Cal Failed (Calibration failed)

Probable Cause

Recommended Corrective Action

Calibration gas supply low or gas connection
leaking

Check calibration gas supplies and connections. Adjust gas pressure and flow.
Replenish low calibration gas supplies and tighten or repair loose or leaking
connections. When calibration gas supplies are adequate, recalibrate.

O

2

cell degraded or failed

Check O

2

cell impedance by reading the O2 Snsr R value via HART (see Figure 4-3,

sheet 2). If cell impedance is zero, replace O

2

cell. If cell impedance is less than 5000

ohms, check for cell housing ground fault. Repair ground fault. If cell impedance is
greater than 5000 ohms and no ground fault is indicated, replace O

2

cell.

Flow calibration gas to the O

2

cell. Read the cell millivolt output. Plot the cell millivolt

output and the calibration gas O

2

concentration on the chart shown in Figure 7-2. If the

plotted values do not fall on the slope line shown in Figure 7-2, replace the O

2

cell.

COe sensor degraded

Replace COe sensor or increase warning level (SYSTEM, CALIB SETUP, COe Slope
Warn
via HART.)

Low sample gas flow in sensor housing due to
flow path plugging

Check the following portions of the flow path for plugging:
• blowback filter
• in-situ filter
• eductor outlet path