Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 79

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

6-17

OCX 8800

Disassemble COe Sensor Assembly

1. Carefully remove screws (1, Figure 6-13), lockwashers (2), and COe

sensor (3) from sensor holder (4). Remove and discard gasket (5).

2. If damaged, use the following procedure to remove thermocouple

adaptor (6) from sensor holder (4):
a. Use a propane torch to heat the thermocouple adaptor to 450°F

(232°C), minimum.

b. While heating, use a flat-head screwdriver to apply removal torque.

Apply torque until the pipe thread sealant softens. Remove and
discard the thermocouple adaptor.

c. Use MEK or methylene chloride solvent to clean thread sealant

residue from the internal pipe threads in the sensor holder. Refer to
applicable MSDS sheet for solvent handling precautions.

Figure 6-13. COe Sensor,

Exploded View

37390030

8

9

4

5

6

1

2

3

7

1. Screw
2. Lockwasher
3. COe Sensor
4. Sensor Holder
5. Gasket
6. Thermocouple Adapter
7. Pre-Heater
8. Plug
9. Stainless Steel Balls