Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

4-2

Figure 4-1. Signal Line Connections,

250 Ohms Load Resistance

Method 1, for Load Resistance

250 Ohms

Refer to Figure 4-1 and the following steps to connect the HART
communicator to a signal line 250 ohms or more of load resistance.

Using the supplied lead set, connect the HART communicator in parallel to
the OCX 8800. Use any wiring termination points in the oxygen analog output
4-20 mA signal line.

Loop

Connectors

O 4-20 mA Signal Line

2

HART

Communicator

Model 275/375

RL 250

Analog Output Device

HART Communicator

Rear Panel

Lead Set

37390023

4-20 mA Terminal Block in

Electronics Housing

A OUT 1 -

A OUT 1 +

1

Do not make connections to the HART communicator's serial port, 4-20 mA signal lines, or
NiCad recharger jack in an explosive atmosphere. Explosions can result in death or serious
injury.