Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 87

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

6-25

OCX 8800

Figure 6-20. Installation of O

2

Cell

and Heater Strut Assembly

Install Sample Block Heater Rods

1. Before installing sample block heater rods (13, Figure 6-20), evenly coat

the heater rods with Watlube heater release agent.

5

6

12

11

14

15

10

9

1

2

4

3

7

8

13

3

7

390066

1. Screw
2. Locking Clip
3. Cover
4. O-ring
5. Marking Plate
6. Terminal Insulator
7. Reference Air Tube
8. Sensor Housing
9. Heater Strut Assembly

10. Gasket
11. Screw
12. Heater Clamp
13. Heater Rod
14. Insulator
15. Thermal Switch