Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 18

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

1-10

NOTE
All static performance characteristics are with operating variables constant. Specifications subject to change
without notice.

Sensors Housing

NEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports

Electronics Housing

NEMA 4, IP66 with fitting and pipe on reference exhaust port to
clean, dry atmosphere, two 3/4-14 NPT conduit ports

Electrical Noise

EN 61326-1, Class A

Certifications

Sensor Housing

Electronics Housing

Line Voltage

Universal 100 to 240 VAC ±10%, 50 to 60 Hz, no switches or
jumpers required, 3/4-14 NPT conduit port

Pollution Degree

2

Over Voltage Category

II

Relative Humidity

5 to 95% (non-condensing)

Isolated Output

Oxygen

4-20 mAdc, 950 ohm maximum, with HART capability

Combustibles

4-20 mAdc, 950 ohm maximum

Alarm

Alarm output relay - dry contact, form C, 30 mA, 30VDC capacity

Power Consumption

750 W maximum

Specifications

KEMA 04 ATEX 2308

EEx d IIB+H2 T3

0344

II 2 G

2005.1602514

CLASS 1, ZONE 1

Ex d IIB+H2 T3

AEx D IIB+H2 T3

TYPE 4 / IP66

APPROVED

CLASS 1, ZONE 1
AEx D IIB+H2 T3

TYPE 4 / IP66

C

US

2005.1602514

CLASS 1, ZONE 1

Ex d IIB+H2 T6

AEx D IIB+H2 T6

TYPE 4 / IP66

KEMA 04 ATEX 2308

EEx d IIB+H2 T6

IP66

0344

II 2 G

CLASS 1, ZONE 1
AEx D IIB+H2 T3

TYPE 4 / IP66

APPROVED

C

US