Ocx 8800, Instruction manual, Ocx with remote electronics – Emerson ROSEMOUNT OCX 8800 User Manual

Page 64

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

6-2

OCX with
Remote Electronics

Remove Sensor Housing

1. Turn off power to the system.
2. Shut off the test gasses at the cylinders and close the instrument air

valve.

3. Disconnect the calibration gas, reference air, eductor air, and dilution air

lines from the sensor housing, Figure 6-1.

Figure 6-1. OCX with

Remote Electronics

Heater

Power Cable

Low O Test Gas

2

High O Test Gas

2

CO Test Gas

Instrument Air

(Reference Gas)

Test

Gas

Ref

erence

Air

Dilution

Air

Eductor Air

Instrument Air

Duct

Stack

AC Power

Input

Signal Cable

37390044

Sensor

Housing

Electronics
Housing

4-20 mA

Outputs