1. Verify that the replacement tube fitting is identical to the item removed.

Special fittings are etched with code markings.

2. Apply pipe thread sealant (Loctite #567) to the mating threads of the

tube fitting. Do not apply sealant to the first turn of the external pipe
threads.

3. Install and tighten the tube fitting in the mating port of sensor or

electronics housing.