SET TEST GAS VALUES

Use HART/AMS to set test gas values for calibration. Refer to Section 4,
Using HART Communications for more information.

Setting Test Gas Values with HART

1. Use the HART communicator or AMS software to access the HART

menu.

2. From the DETAILED SETUP menu, select O2 CALIB PARAMS.
3. From O2 CALIB PARAMS, select O2 High Gas. Enter the percent O

2

used for the high O

2

test gas.

4. From O2 CALIB PARAMS, select Low TG. Enter the percent O

2

used

for the low O

2

test gas.

5. From the DETAILED SETUP menu, select COe CALIB PARAMS.
6. From COe CALIB PARAMS, select COe Test Gas. Enter the CO

concentration (ppm) used for COe test gas.

OCX 8800 RESET
PROCEDURE

Whenever you correct an equipment alarm or fault condition, the OCX 8800
will either revert to normal operation or continue to indicate an alarm status
condition. If the equipment does not revert to normal operation when a fault
condition is cleared, or if instructed to do so in Section 7, Troubleshooting,
use the following procedure to reset the OCX 8800.

OCX Reset with HART

Remove the OCX 8800 from the process loop and recycle power.