Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

7-6

ADC Failure (Voltage to digital conversion could not complete)
ADC Ref Error (Voltage to digital conversion not accurate)

Probable Cause

Recommended Corrective Action

Incorrect wiring between electronics and sensor
housings

Check all wiring between the electronics and sensor housings per Figure 7-3. Correct
wiring faults.

Electronics package failure

Replace electronics package.

O2 Htr Failure (Oxygen sensor heater could not reach final temperature)

Probable Cause

Recommended Corrective Action

O

2

heater circuit wiring open

Check O

2

cell heater circuit for broken wire or loose connection per Figure 7-3 and

Figure 6-6. Repair broken wire or loose connection.

O

2

heater open

Check resistance of O

2

heater per Figure 6-6. Normal O

2

heater resistance is 62.5

Ohms. Replace O

2

heater if heater is open or has a large resistance.

Heater electronics failure

Check heater fuse F3 in electronics housing per Figure 7-1. If open, locate and correct
cause of overload. If F3 is not open, or if cause of overload cannot be found, replace
electronics package.

COe Htr Failure (Combustibles sensor heater could not reach final temperature)

Probable Cause

Recommended Corrective Action

COe heater circuit wiring open

Check COe heater circuit for broken wire or loose connection per Figure 7-3 and
Figure 6-7. Repair broken wire or loose connection.

COe heater open

Check resistance of COe heater per Figure 6-7. Normal COe heater resistance is 97.7
Ohms. Replace COe heater if heater is open or has a large resistance.

Heater electronics failure

Check heater fuse F3 in electronics housing per Figure 7-1. If open, locate and correct
cause of overload. If F3 is not open, or if cause of overload cannot be found, replace
electronics package.

SB Htr Failure (Sample block heater could not reach final temperature)

Probable Cause

Recommended Corrective Action

Sample block heater circuit wiring open

Check sample block heater circuit for broken wire or loose connection per Figure 7-3
and Figure 6-6. Repair broken wire or loose connection.

Sample block heater open

Check resistance of sample block heater per Figure 6-6. Normal sample block heater
resistance is 36.4 Ohms each (18.2 Ohms with both heaters in parallel). Replace
sample block heater if heater is open or has a large resistance.

Heater electronics failure

Check heater fuse F4 in electronics housing per Figure 7-1. If open, locate and correct
cause of overload. If F4 is not open, or if cause of overload cannot be found, replace
electronics package.

Sensor housing exposed to high wind and/or
extreme cold temperatures

If above probable causes are not causing the SB heater failure, install flange insulator
(PN 6P00162H01).

Table continued on next page