Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

2-16

Figure 2-9. Pneumatic

Installation, OCX with Reference

Air Set and Solenoids (with

Autocalibration)

Reference Air Set and Solenoids Option

When the reference air set and test gas solenoids are included with your
OCX 8800, use the following procedure to install the pneumatic system
components.

1. Install the reference air set according to the instructions in Reference Air

Set Option, steps 1 through 3.

2. Refer to Figure 2-9. Connect the O

2

low gas source to the CAL GAS LO

O

2

inlet fitting on the electronics housing. Install a shutoff valve and

pressure regulator with gage in the O

2

low supply line, as shown.

3. Connect the O

2

high gas source to the CAL GAS HI O2 inlet fitting.

Install a shutoff valve and pressure regulator with gage in the O

2

high

supply line.

Electronics

Housing

Dilution

Air In

Reference

Air In

CAL Gas In

CAL Gas

Flow Meter

7 scfh, 20-30 psig

Recommended

Dilution Air

Flow Meter

0.1 scfh

Instrument

Air Supply

Pressure Regulator/Filter

35 psig - General Purpose

45 psig - Hazardous Area

2-Stage

Regulators

CO

HI O

2

LO O

2

CAL Gas Out

Eductor

Air In

Sensor

Housing

Instrument Air Out

(

(

37390012