Ocx 8800 – Emerson ROSEMOUNT OCX 8800 User Manual

Page 69

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

6-7

OCX 8800

Install Remote Electronics Housing

1. Mount remote electronics housing on wall or pipe within distance of

signal and heater cables in use.

2. Remove the electronics housing cover.
3. If removed, install the power and signal cables and the customer power

and signal conduits and wiring at the electronics housing.

4. Connect the signal cable leads to the O

2

cell and thermocouple

connector (J4), and to the COe and CJC connector (J5), Figure 6-4.

5. Connect the heater cable leads to the heater power connector (J3).
6. Connect the line (L1 wire) to the L1 terminal, and the neutral (N wire) to

the N terminal on the AC power input terminal block, Figure 6-4.

7. Connect the ground lead to the ground stud. Secure the connection with

two nuts. Attach a separate ground lead (G wire) from the ground stud
to the G terminal on the power input terminal block.

8. If used, connect external relay leads to the alarm output relay terminal.
9. Connect the test gas and instrument air lines to the remote electronics

housing. Connect the calibration gas line and instrument air line to the
remote electronics housing.

10. Refer to Figure 6-1 and Figure 6-4. Make sure all test gas lines and

electrical connections are complete.

11. Install the cover on the electronics housing.

12. Turn on the test gasses at the cylinders and open the instrument air

valve.

13. Restore power to the system.

Advertising