Ocx 8800 – Emerson ROSEMOUNT OCX 8800 User Manual

Page 76

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

6-14

Remove Eductor

The O

2

cell and heater strut assembly (9, Figure 6-5) and the COe sensor

assembly (5, Figure 6-8) must be removed before you start this procedure.

1. Use straightedge to matchmark alignment of eductor flat and elbow, as

shown in Figure 6-10.

Figure 6-10. Eductor Alignment

Matchmarks

2. Unscrew terminal block mounting (13, Figure 6-8). Move terminal block

mounting away from eductor (6).

3. Unscrew eductor holder (14) with eductor (6) and fittings (12 and 15)

from sensor housing (7).

4. Clamp flats of eductor (6) in jaws of bench vise.

5. Use a propane torch to heat the eductor (6) to 450°F (232°C), minimum.

Apply the heat near the screw threads.

6. While heating the eductor (6), use wrench to apply removal torque to

elbow (12), eductor holder (14), or tube fitting (15) until the thread
sealant softens. Remove mating parts (12, 14, and 15).

7. Use MEK or methylene chloride solvent to clean thread sealant residue

from the pipe threads of the mating parts. Refer to applicable MSDS
sheet for solvent handling precautions.

37390036

Eductor Flat

Matchmark

Straightedge

Use heat resistant gloves when removing mating parts from the eductor. The mating parts
are bonded with a thread sealing compound. The compound softens at 450°F (232°C). The
heated parts can cause severe burns.

Advertising