Instruction Manual

IM-106-880, Rev 1.0
January 2007

6-11

OCX 8800

2. Loosen screws (11, Figure 6-5) and rotate heater clamps (12) to release

heater rods (13). One heater clamp secures each heater rod.

3. Slide sample block heater rods (13) out of housing (8).
4. To replace thermal switch (15), remove insulator (14). Disconnect

heater wires. Unscrew and remove thermal switch.

Remove COe Sensor Assembly

1. Disconnect COe heater, thermocouple, and sensor wires from terminal

blocks. Refer to Figure 6-7.

2. Remove insulator (1, Figure 6-8).

Figure 6-7. COe Sensor,

Thermocouple, and Heater

Connections

O2

RED

T/C O2

T/C CO

T/C SB

EXC

CJC

COe Sensor Wires

COe Heater Wires

COe Sensor

Assembly

CO REF

CO ACT

Sensor Housing

T

erminals

+

-

HTR 02

HTR SB

2

2

1

+

-

+

-

+

-

+

-

+

-

+

YEL

-

+

37390018

HTR 02

2

1

HTR CO

1

WHT

BLU

RED

RED

WHT

COe Thermocouple Wires

- EXC

RED

BLU

CJC Sensor

NOTE: All wires
at these terminals
are in the CJC
current loop.