background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

6-3

OCX 8800

4. Remove the cover from the sensor housing to expose the sensor

housing terminal blocks, Figure 6-2.

5. Disconnect the signal cable from the O

2

and T/C terminal blocks, and

from the CO and CJC terminal blocks.

6. Disconnect the heater power cable from the HTR terminal blocks.
7. If moving the sensor housing to another work site, disconnect and

remove the power and signal cables from the sensor housing.

8. Remove insulation to access the mounting bolts. Unbolt the sensor

housing from the stack and take it to a clean work area.

9. Allow the sensor housing to cool to a comfortable working temperature.

Figure 6-2. Sensor

Housing Terminals

3

7390028

BLU

ORG

YEL

RED

BLK

WHT

BLK

YEL

RED

BLU

WHT

BLK

BLK

GRN

BLK

BLK

BLK

ORG

WHT

RED

BLK

BRN

To

Ground

Screw

GRN

SENSOR HOUSING

Signal Cable

Heater Power Cable

T/C O2

T/C CO

T/C SB

EXC

CJC

CO REF

CO ACT

+

-

+

-

+

-

+

-

+

-

+

-

+

EXC

-

O2

HTR SB

2

2

1

-

+

HTR 02

2

1

HTR CO

1

NOTE: Wire colors shown are for cables

supplied by Emerson Process
Management.