Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 156

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

B-4

NOTE
The O

2

signal connection is not required for SPA configuration or calibration.

The OCX interface will allow you to observe the O

2

signal level when the SPA

configuration procedure is completed.

4. Observe the front panel of the SPA, Figure B-5:

a. A process value display in the SPA display window indicates that the

SPA is operational. Four pushbuttons are located below the display
window.

b. Pressing a left-hand pushbutton scrolls up ( ) or down ( ) through

the SPA command menu, a submenu, or parameter values list.

c. Pressing the VIEW pushbutton displays rail limits and alarm relay

configurations. There are five sequential displays in the VIEW mode.
While in the VIEW mode, the up ( ), down ( ), and SELECT
pushbuttons are disabled.

NOTE
In the VIEW mode, you can scroll through and display the output zero and full
scale settings and the alarm relay trip points and configurations.

d. Pressing the SELECT pushbutton selects the displayed menu or

submenu command or selects a displayed parameter variable.

Figure B-5. SPA Front Panel

37020017

SPA

Display

Window

Left-Hand

Pushbuttons

Right-Hand

Pushbuttons