Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 155

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0
January 2007

B-3

OCX 8800

IN

IN

Fluke Model 87

Multimeter

or equivalent

Optional

mA

AC or DC

Power Input

AC

ACC

GND

37390024

250

950

>
<

OCX
8800

Figure B-3. SPA Jumper

and Dip Switch Settings

Figure B-4. SPA Setup for

Calibration

2. Verify that the Password Jumper is set to the OFF position. If the jumper

is in the ON position, reposition the jumper.

3. Check the position of the Failsafe Dip Switches. Position the dip

switches as shown in Figure B-3.

4. Check the position of the Source Current Dip Switches. Position the dip

switches as shown in Figure B-3.

5. Reinstall the SPA access cover.

Configuration/Calibration

Prior to operation, the SPA operating parameters must be configured via a
menu-driven setup procedure. At the end of the configuration procedure, the
SPA analog output signal is calibrated to insure valid communications.

1. See Figure B-4. Connect a calibrated ammeter (Fluke Model 87 or

equivalent, accurate to ±0.025%) to the SPA analog output terminals.
Observe polarity.

2. Connect a 90 to 260 VAC or 22 to 300 VDC power source to the SPA

power terminals. When connecting an AC power source, use the AC
and ACC (AC Common) terminals. For a DC source, use the AC and
Ground terminals.

3. If desired, you can connect the 4 to 20 mA O2 signal wires from the

OCX 8800 analog output terminal block to the SPA Input terminals. (The
OCX must be operational to transmit the O2 signal. Observe polarity.)

Access Cover

37020022

Password

ON

Password

OFF

Password

Jumper Pins

Failsafe Dip Switches

(shown in correct position)

(shown in correct position)

Source Current Dip Switches

SPA - Bottom View

NOTICE

REMOVE THIS PANEL

FOR SERVICE ACCESS.

STATIC SENSITIVE.

USE CAUTION WHEN

CHANGING JUMPERS

Advertising