Ocx 8800, Instruction manual – Emerson ROSEMOUNT OCX 8800 User Manual

Page 72

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

6-10

Figure 6-6. O

2

Cell,

Thermocouple, and

Heater Connections

Remove O

2

Cell and Heater Strut Assembly

1. Remove reference air tube (7, Figure 6-5) from sensor housing (8).
2. See Figure 6-6. Disconnect and tag O

2

heater wires, O

2

cell and return

wires, and thermocouple wires at the sensor housing terminals.

3. Remove the O

2

cell and heater strut assembly (9, Figure 6-5) from

sensor housing (8). Remove and discard gasket (10).

Remove Sample Block Heater Rods

1. Disconnect sample block heater rod wires from terminal block. Refer to

Figure 6-6.

+

+

+

+

O2

Thermocouple Wires

ORG

RED

T/C O2

T/C CO

T/C SB

EXC

CJC

O Cell Wires

2

O Heater Wires

2

Return

Wire

O Cell and Heater Strut

Assembly

2

CO REF

CO ACT

Sensor Housing

T

erminals

+

-

HTR 02

HTR SB

2

GRN

2

1

+

-

+

-

+

-

+

-

+

-

+

YEL

-

+

+

37390069

HTR 02

2

1

Sample Block

Heater Rods

Thermal

Switch

HTR CO

1

EXC

-

Sample

Block

Thermo-

couple

Advertising