Ocx 8800 – Emerson ROSEMOUNT OCX 8800 User Manual

Page 54

Advertising
background image

Instruction Manual

IM-106-880, Rev 1.0

January 2007

OCX 8800

4-10

COe D/A trim procedure using HART/AMS

Use the following procedure to perform the COe D/A trim procedure using the
HART communicator or AMS. Refer to the HART menu tree, Figure 4-3.

NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.

To return to a preceding menu, press the left arrow key.

1. From the DIAG/SERVICE menu, select D/A TRIM. The HART

communicator displays O2 D/A Trim. Press the up or down arrow to
select COe D/A Trim.

2. Press the right arrow key to start the procedure. (If you wish to exit D/A

Trim with no changes, select ABORT.)

3. The HART communicator displays WARNING: Loop should be

removed from automatic control. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press OK.

4. Remove the electronics housing cover.
5. Refer to Figure 2-5. Connect a digital multimeter to read the milliamp

output from the COe D/A coverter circuit. Connect the positive lead to
the AOUT2+ terminal and connect the negative lead to the AOUT2-
terminal. Then, press OK at the HART communicator.

6. The HART communicator displays Setting Fld dev output to 4 mA.

Press OK. Read the COe

millamp output at the digital multimeter. Enter

the reading at the HART communicator and press ENTER. (Select
ABORT to exit without changes).

7. The HART communicator displays Setting Fld dev output to 20 mA.

Press OK. Read the COe

millamp output at the digital multimeter. Enter

the reading at the HART communicator and press ENTER. (Select
ABORT to exit without changes).

8. The HART communicator displays Setting Fld dev output to 4 mA.

Press OK.

9. The HART communicator displays Fld dev output 4.00 mA equal to

reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 6.

10. The HART communicator displays Setting Fld dev output to 20 mA.

Press OK.

11. The HART communicator displays Fld dev output 20.00 mA equal to

reference meter? Using the up or down arrow, select 1 Yes or 2 No
and Press ENTER. If No, the process repeats from step 7.

12. The HART communicator displays NOTE: Loop may be returned to

automatic control.