Exatron 3000b – Extron electronic 3000B User Manual

Page 127

Advertising
background image

Exatron 3000B

www.exatron.com

7-2

Chapter 7 Preventive Maintenance

5. Wipe down all stainless steel surfaces (non-trackwork) with C

HAMPION

brand Stainless Steel

Cleaner (part #TLS09-125) sprayed onto a clean cloth. DO NOT SPRAY STAINLESS STEEL
CLEANER DIRECTLY ONTO THE HANDLER. DO NOT USE STAINLESS STEEL CLEANER ON
HANDLER TRACKWORK.

6. Clean the filter on the vacuum generator (part #VGH10F-1/4-1/4.) This filter (part #VGFE10) is white

when clean, and is easily visible through the clear plastic cover over the vacuum generator.

VACUUM GENERATOR MAINTENANCE IS DISCUSSED IN GREATER DETAIL AND A CALL

-

OUT ILLUSTRATION IS

PROVIDED IN THE

"V

ACUUM

G

ENERATOR

M

AINTENANCE

G

UIDELINES

"

SECTION LATER IN THIS CHAPTER

.

DIRTY FILTER

CLEAN FILTER

A. Disconnect the flexible plastic air tube from the swivel elbow where it attaches to the

plunger assembly, by pulling the air tube straight out.

B. Loosen the vacuum generator's gray thumb screw which holds the filter chamber in place.

Remove the filter and clean it with water, or better yet, replace the filter element with a
new one. Replace/reinstall filter and tighten the gray thumb screw finger tight only.

7.

Upon Power Up, or As Needed:

Check octoloader alignment. To do so, press the [0] key on the handler controller to turn OFF all
motors. The handler display will read, ALL MOTORS OFF. Manually move the octoloader plate all
the way to the right until it stops.

At this point, the left octoloader track should be centered above the test site area. If all is well, press
the [Enter] button on the handler controller. The handler display will read ALIGN MOTORS ? Press
the [Enter] button again to resume operation.

If the left octoloader track is not centered above the test site area, press the [Enter] button on the
handler controller once. This will turn the motors back on and the handler display will read,
ALIGN MOTORS ?

The operator may now "jog" the octoloader by one .003" step at a time until it is properly aligned, using
the keypad keys as described below:

[1] Jog the octoloader left, display will read, JOG LEFT

[3] Jog the octoloader right, display will read, JOG RIGHT

After alignment is complete, press the [Enter] button again to resume operation. This entire
octoloader alignment procedure may be repeated as necessary. If you have any trouble with this
procedure, call the E

XATRON

factory for assistance.

Our toll-free telephone number is 1-800-EXA-TRON.

Advertising