Dynamic characteristics – National Instruments AT E Series User Manual

Page 109

Advertising
background image

Appendix A

Specifications for AT-MIO-16E-1, AT-MIO-16E-2, and AT-MIO-64E-3

AT E Series User Manual

A-4

ni.com

Dynamic Characteristics

Bandwidth

Settling time for full-scale step

Small

Signal (–3dB)

Large Signal

(1% THD)

AT-MIO-16E-1

1.6 MHz

1 MHz

AT-MIO-16E-2,
AT-MIO-64E-3

1 MHz

300 kHz

Gain

Accuracy*

±0.012%

(±0.5 LSB)

±0.024%

(±1 LSB)

±0.098%
(±4 LSB)

AT-MIO-16E-1

0.5

2 µs typ
3 µs max

1.5 µs typ
2 µs max

1.5 µs typ
2 µs max

1

2 µs typ
3 µs max

1.5 µs typ
2 µs max

1.3 µs typ
1.5 µs max

2 to 50

2 µs typ
3 µs max

1.5 µs typ
2 µs max

0.9 µs typ
1 µs max

100

2 µs typ
3 µs max

1.5 µs typ
2 µs max

1 µs typ
1.5 µs max

AT-MIO-16E-2

All

2 µs typ
4 µs max

1.9 µs typ
2 µs max

1.8 µs typ
2 µs max

AT-MIO-64E-3

All

3 µs typ
5 µs max

2 µs typ
3 µs max

1.8 µs typ
2 µs max

* Accuracy values valid for source impedances < 1 k

Ω. Refer to the

Multiple-Channel

Scanning Considerations

section of Chapter 3,

Hardware Overview

, for more information.

Advertising