background image

Appendix A

Specifications for AT-MIO-16XE-50

AT E Series User Manual

A-36

ni.com

Data transfers ..........................................DMA, interrupts,

programmed I/O

DMA modes ...........................................Single transfer

Triggers

Digital Trigger

Compatibility ..........................................TTL

Response .................................................Rising or falling edge

Pulse width .............................................10 ns min

RTSI

Trigger Lines ..........................................7

Calibration

Recommended warm-up time.................15 min

Calibration interval .................................1 year

External calibration reference.................>6 and <9.999V

Onboard calibration reference

Level ................................................5.000 V (±3 mV) (over full

operating temperature, actual
value stored in EEPROM)

Temperature coefficient...................±2 ppm/°C max

Long-term stability ..........................±15 ppm/

Bus Interface

Type ........................................................Slave

Power Requirement

+5 VDC (±5%) .......................................0.75 A

Power available at I/O connector............+4.65 VDC to +5.25 VDC

at 1 A

1,000 h