Chapter 3

Hardware Overview

AT E Series User Manual

3-8

ni.com

AT-MIO-16E-10, and AT-MIO-16DE-10 have gains of 0.5, 1, 2, 5, 10, 20,
50, and 100 and are suited for a wide variety of signal levels. With the
proper gain setting, you can use the full resolution of the ADC to measure
the input signal. Table 3-2 shows the overall input range and precision
according to the input range configuration and gain used.

♦ AT-MIO-16XE-10, AT-AI-16XE-10, AT-MIO-16XE-50

These devices have two input polarities—unipolar and bipolar. Unipolar
input means that the input voltage range is between 0 and V

ref

, where V

ref

is a positive reference voltage. Bipolar input means that the input voltage
range is between –V

ref

and +V

ref

. The AT-MIO-16XE-10, AT-AI-16XE-10,

and AT-MIO-16XE-50 have a unipolar input range of 10 V (0 to 10 V) and
a bipolar input range of 20 V (±10 V). You can program polarity and range
settings on a per channel basis so that you can configure each input channel
uniquely.

Table 3-2. Actual Range and Measurement Precision

Range

Configuration

Gain

Actual Input Range

Precision

1

0 to +10 V

1.0
2.0
5.0

10.0
20.0
50.0

100.0

0 to +10 V

0 to +5 V
0 to +2 V
0 to +1 V

0 to +500 mV
0 to +200 mV
0 to +100 mV

2.44 mV
1.22 mV

488.28 µV
244.14 µV
122.07 µV

48.83 µV
24.41 µV

–5 to +5 V

0.5
1.0
2.0
5.0

10.0
20.0
50.0

100.0

–10 to +10 V

–5 to +5 V

–2.5 to +2.5 V

–1 to +1 V

–500 to +500 mV
–250 to +250 mV
–100 to +100 mV

–50 to +50 mV

4.88 mV
2.44 mV
1.22 mV

488.28 µV
244.14 µV
122.07 µV

48.83 µV
24.41 µV

1

The value of 1 LSB of the 12-bit ADC; that is, the voltage increment corresponding to a

change of one count in the ADC 12-bit count.

Note: Refer to Appendix A,

Specifications

, for absolute maximum ratings.