ni.com

Differential Connections for Nonreferenced or
Floating Signal Sources

Figure 4-7 shows how to connect a floating signal source to an AT E Series
device channel configured in DIFF input mode.

Figure 4-7. Differential Input Connections for Nonreferenced Signals

Figure 4-7 shows two bias resistors connected in parallel with the signal
leads of a floating signal source. If you do not use the resistors and the
source is truly floating, the source is not likely to remain within the
common-mode signal range of the PGIA, and the PGIA saturates, causing
erroneous readings. You must reference the source to AIGND. The easiest
way is simply to connect the positive side of the signal to the positive input
of the PGIA and connect the negative side of the signal to AIGND as well

PGIA

Selected Channel in DIFF Configuration

Programmable

Gain

Instrumentation

Amplifier

+

+

Input Multiplexers

AISENSE

AIGND

I/O Connector

+

V

s

ACH+

ACH–

V

m

Bias

Current

Return

Paths

Bias
Resistors
(see text)

Floating

Signal

Source

Measured

Voltage