Calibration, Bus interface, Power requirement – National Instruments AT E Series User Manual

Page 133: Physical, Maximum working voltage

Advertising
background image

Appendix A

Specifications for AT-MIO-16XE-10 and AT-AI-16XE-10

AT E Series User Manual

A-28

ni.com

Calibration

Recommended warm-up time.................15 min

Calibration interval .................................1 year

External calibration reference.................>6 and <9.999V

Onboard calibration reference

Level ................................................5.000 V (±1.0 mV) (over full

operating temperature, actual
value stored in EEPROM)

Temperature coefficient...................±0.6 ppm/°C max

Long-term stability ..........................±6 ppm/

Bus Interface

Type ........................................................Slave

Power Requirement

+5 VDC (±5%) .......................................1.2 A

Power available at I/O connector............+4.65 VDC to +5.25 VDC

at 1 A

Physical

Dimensions
(not including connectors) ......................33.8 by 9.9 cm

(13.3 by 3.9 in)

I/O connector ..........................................68-pin male SCSI-II type

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the
common-mode voltage.

Channel-to-earth .....................................42 V, Installation Category II

Channel-to-channel.................................42 V, Installation Category II

1,000 h

Advertising