Pulse modulation rf level frequency response, Test procedure, Pulse modulation on/off ratio – VXI 3002 User Manual

Page 128

Advertising
background image

Annex-A-2

Pulse modulation RF level frequency response

Fig. A-1 Pulse modulation test set-up

Test procedure

(1)

Perform AUTO ZERO and AUTO CAL on the power meter.

(2)

Connect the test equipment as shown in Fig. A-1.

(3)

On the UUT set:

Carr Freq

100 kHz

RF

Level

−7 dBm

Pulse

ON

(4)

Set the function generator to provide +5 V DC. The RF output will now be enabled.

(5)

Record the output level measured by the power meter against each of the carrier frequencies
shown in Table A-1, checking that the results are within specification.

(6)

Set the UUT RF level to +0 dBm and repeat (5) using Table A-2.

(7)

Set the UUT RF level to +16 dBm and repeat (5) using Table A-3.

Pulse modulation on/off ratio

C5358

OUTPUT

Function generator

RF

OUT

UUT

PULSE

INPUT

RF INPUT

Spectrum analyzer

Fig. A-2 Pulse modulation on/off ratio test set-up