Internal fm, Fm deviation and distortion – VXI 3002 User Manual

Page 85

Advertising
background image

5-10

Internal FM

Specification

Deviation range

0 to 100 kHz.

Resolution

3 digits or 1 Hz.

Accuracy

±5% at 1 kHz modulation rate.

Bandwidth (1 dB)

DC to 100 kHz (DC coupled),
10 Hz to 100 kHz (AC coupled),
20 Hz to 100 kHz (AC coupled with ALC).

Carrier frequency offset

Less than 1% of the set frequency deviation when DC coupled.

Distortion

Less than 3% at 1 kHz rate for deviations up to 100 kHz.
Typically <0.5% at 1 kHz rate for deviations up to 10 kHz.

FSK

2 level or 4 level FSK.
Note that 4 FSK is not available with Option 11 Fast Pulse fitted.

Test equipment

Description Minimum

specification Example

Modulation meter

FM accuracy

±1% at 1 kHz modulation frequency

IFR 2305 with
distortion option

DVM

DC voltage measurement

Solartron 7150+

50

Ω load

(termination)

1 W, 50

Ω nominal impedance, DC to 2.4 GHz

Lucas Weinschel
M1404N

Audio analyzer

Capable of measuring THD of 0.01% from 100 Hz
to 20 kHz

Rohde &
Schwarz UPA3

Function
generator

DC to 100 kHz sine,

±0.6 dB flatness

Agilent 3325B

FM deviation and distortion

Test procedure

UUT

2305

Modulation meter

C2564

RF INPUT

RF OUT

Fig. 5-7 Internal modulation and modulation distortion test set-up