:UNITS

Select default RF level units.

Data type : Character Program Data (DBM, DBV, DBMV, DBUV, V, MV or UV)

Allowed suffices : None

Default suffix : None

Examples: RFLV:VALUE -27.3DBM;ON

RFLV:TYPE PD;VALUE 1.23UV

:LIMIT

Set RF Level max limit (short form)

:VALUE

Set RF Level max limit

Data type : Decimal Numeric Program Data

Allowed suffices : Any one of: DBM, DBV, DBMV, DBUV, V, MV or UV

Default suffix : DBM unless changed by UNITS command

:ENABLE

Enable

limit

:DISABLE

Disable

limit

:OFFS

Set RF Level offset for given frequency band (short form)

:VALUE

Set RF Level offset for given frequency band

Data type : Decimal Numeric Program Data

Allowed suffices : DB only

Default suffix : DB

:ENABLE

Enable

offsets

:DISABLE

Disable

offsets

:SAVE

Save offsets in non-volatile memory

RFLV?

Prepares message containing information on RF Level setting in the
following format:

:RFLV:UNITS <unit>;TYPE <type>;VALUE <nr2>;INC <nr2>;<status>

where:

<unit> is character program data defining the default RF
level units (DBM, DBV, DBMV, DBUV, V, MV or UV),
<type> is character program data indicating EMF or PD and
<status> is a program mnemonic indicating whether the RF
output is ON or OFF

Examples: :RFLV:UNITS DBM;TYPE PD;VALUE -103.5;INC 2.0;ON

:RFLV:UNITS DBV;TYPE EMF;VALUE -83.2;INC 0.5;ON

RFLV:LIMIT?

Prepares message containing information on RF Level max limit setting
in the following format:

:RFLV:LIMIT:VALUE<nr2>;<status>;

Examples: :RFLV:LIMIT:VALUE-20.0;ENABLE

RFLV:OFFS?

Prepares message containing information on RF Level offset in the
following format:

:RFLV:OFFS:VALUE<nr2>;<status>;

Examples: :RFLV:OFFS:VALUE-3.2;ENABLE