Rf level frequency response, Alc linearity – VXI 3002 User Manual

Page 79

Advertising
background image

5-4

RF level frequency response

Test procedure

UUT

C2558

RF OUT

SENSOR

INPUT

6960B

RF power meter

6932

Power

sensor

Fig. 5-1 RF output test set-up

(1) Perform AUTO ZERO and AUTO CAL on the power meter.

(2) Connect the test equipment as shown in Fig. 5-1.

(3) On the UUT set:

Carr Freq

30 kHz

RF Level

0 dBm

(4) Record the output level measured by the power meter against each of the carrier frequencies

shown in Table 5-1 checking that the results are within specification.

(5) Set the UUT RF level to +7 dBm and repeat (4) using Table 5-2.

(6) Set the UUT RF level to +25 dBm and repeat (4) using Table 5-3, decreasing the RF level to

+19 dBm when testing at carrier frequencies above 1.2 GHz.

ALC linearity

Test procedure

(1) Perform AUTO ZERO and AUTO CAL on the power meter.

(2) Connect the test equipment as shown in Fig. 5-1.

(3) On the UUT set:

Carr Freq

2.5 MHz

RF

Level

−4 dBm

(4) Record the output level measured by the power meter against each of the steps shown in

Table 5-4 checking that the results are within specification.

(5) Set the UUT carrier frequency to 950 MHz and repeat (4) using Table 5-5.

(6) Set the UUT carrier frequency to 1200 MHz and repeat (4) using Table 5-6.

(7) Set the UUT carrier frequency to 1900 MHz and repeat (4) using Table 5-7.

(8) Set the UUT carrier frequency to 2400 MHz and repeat (4) using Table 5-8.

Advertising