Carrier frequency – VXI 3002 User Manual

Page 36

Advertising
background image

3-6

Carrier frequency

These commands enable you to set the carrier frequency in the range 9 kHz to 2.4 GHz to a
resolution of 1 Hz. You can adjust the frequency in steps by setting the size of the step and then
stepping the frequency up or down. After having adjusted the frequency you can either return to
the reference frequency or make the current frequency the reference frequency. Additionally, you
can adjust the phase offset of the carrier in degrees in the range -359.99° to +359.99°. Also you
can configure the instrument as a swept frequency signal generator where you define the start and
stop frequencies and set the step size, step time and step direction. For triggering methods, refer to
'Trigger source' at the end of this section.

CFRQ

Set Carrier Frequency (short form)

:VALUE

Set Carrier Frequency

:INC

Set Carrier Frequency step size

Data type : Decimal Numeric Program Data

Allowed suffices : Any one of: GHZ, MHZ, KHZ or HZ

Default suffix : HZ

:UP

Go UP one step

:DN

Go DOWN one step

:RETN

Return to original setting

:XFER

Transfer current value to be the new setting

Data type : None

Allowed suffices : None

Default suffix : None

:MODE

Selects the mode of carrier frequency operation. SWEPT enables swept
carrier frequency operation, while FIXED disables it

Data type : Character Program Data (FIXED - non swept mode, SWEPT - swept

mode)

Allowed suffices : None

Default suffix : None

:START

Set Start Frequency for use in sweep

:STOP

Set Stop Frequency for use in sweep

Data type : Decimal Numeric Program Data

Allowed suffices : Any one of: GHZ, MHZ, KHZ or HZ

Default suffix : HZ

:TIME

Set time per sweep step

Data type : Decimal Numeric Program Data

Allowed suffices : MS or S

Default suffix : MS

:PHASE

Adjust Phase Offset of Carrier in degrees

Data type : Decimal Numeric Program Data

Allowed suffices : DEG

Default suffix : DEG

Examples: CFRQ:VALUE 2.54MHZ;INC 10KHZ

CFRQ:UP;XFER
CFRQ:START 1MHZ;STOP 10MHZ;TIME 100MS
CFRQ:MODE SWEPT

CFRQ?

Prepares message containing information on Carrier Frequency setting
in the following format:

:CFRQ:VALUE <nr2>; INC <nr2>;MODE<mode>

where:

<mode> is character program data indicating whether carrier
frequency operation is swept or fixed

Example: :CFRQ:VALUE 1000000000.0;INC 25000.0;MODE FIXED

Advertising