Rf output, Test procedures, Specification – VXI 3002 User Manual

Page 78: Test equipment

Advertising
background image

5-3

TEST PROCEDURES

Each test procedure relies on the UUT being set to its power-up conditions. Reset the UUT after
each test procedure by setting:

Store/Recall Address 999

Recall

At the end of this chapter are a set of results tables which give all the test points for each of the
tests. These tables should be photocopied and used to record the results of all the measurements
taken.

RF output

Specification

Range

−137 dBm to +25 dBm (+19 dBm above 1.2 GHz).

When AM is selected, the maximum RF output level decreases linearly with increasing
AM depths to +19 dBm (+13 dBm above 1.2 GHz) at 99% depth.

Accuracy

VSWR

For output levels less than

−5 dBm output VSWR is less than 1.3:1 for carrier

frequencies up to 1.2 GHz and less than 1.5:1 for carrier frequencies up to 2.4 GHz.

Output impedance

50

Ω SMA female connector to MIL 390123D.

Test equipment

Description Minimum

specification Example

Power meter

±0.1 dB from 9 kHz to 2.4 GHz

IFR 6960B and
6932

Measuring
receiver

0 dBm to

−127 dBm; 2.5 MHz to 2.4 GHz

Agilent 8902A
with 11722A
sensor and
11793A down
converter

Signal generator

+8 dBm from 32.5 MHz to 2.43 GHz

IFR 2041

Accuracy over temperature range 17

°C to 27°C

9 kHz to 1.2 GHz

1.2 GHz to 2.4 GHz

>

−127 dBm

±1.0 dB

±2.0 dB

Temperature coeff. over temperature range 0

°C to 55°C

9 kHz to 1.2 GHz

1.2 GHz to 2.4 GHz

<±0.02 dB/°C

<±0.04 dB/°C